Abstract
A capillary zone electrophoretic (CZE) method has been developed for the quantitation of complex heroin seized by law enforcement. Comparison with a currently used high-performance liquid chromatographic (HPLC) method demonstrates that the CZE method has the same order of repeatability, is faster, uses minimal organic solvent, and has a resolving power that is far superior to the HPLC procedure.
Original language | English |
---|---|
Pages (from-to) | 1-6 |
Number of pages | 6 |
Journal | Journal of Chromatographic Science |
Volume | 32 |
Issue number | 1 |
DOIs | |
Publication status | Published - 1 Jan 1994 |
Externally published | Yes |