The reliability of CT numbers as absolute values for diagnostic scanning, dental imaging, and radiation therapy simulation: A narrative review

Yazan Al-Hayek, Kelly Spuur, Xiaoming Zheng, Christopher Hayre, Rob Davidson

Research output: Contribution to journalReview articlepeer-review

1 Citation (Scopus)

Abstract

Background and purpose
The purpose of this review was to examine the reported factors that affect the reliability of Computed Tomography (CT) numbers and their impact on clinical applications in diagnostic scanning, dental imaging, and radiation therapy dose calculation.

Methods
A comprehensive search of the literature was conducted using Medline (PubMed), Google Scholar, and Ovid databases which were searched using the keywords CT number variability, CT number accuracy and uniformity, tube voltage, patient positioning, patient off-centring, and size dependence. A narrative summary was used to compile the findings under the overarching theme.

Discussion
A total of 47 articles were identified to address the aim of this review. There is clear evidence that CT numbers are highly dependent on the energy level applied based on the effective atomic number of the scanned tissue. Furthermore, body size and anatomical location have also indicated an influence on measured CT numbers, especially for high-density materials such as bone tissue and dental implants. Patient off-centring was reported during CT imaging, affecting dose and CT number reliability, which was demonstrated to be dependent on the shaping filter size.

Conclusion
CT number accuracy for all energy levels, body sizes, anatomical locations, and degrees of patient off-centring is observed to be a variable under certain common conditions. This has significant implications for several clinical applications. It is crucial for those involved in CT imaging to understand the limitations of their CT system to ensure radiologists and operators avoid potential pitfalls associated with using CT numbers as absolute values for diagnostic scanning, dental imaging, and radiation therapy dose calculation.
Original languageEnglish
Pages (from-to)138-146
Number of pages9
JournalThe Canadian journal of medical radiation technology / CAMRT
Volume53
Issue number1
DOIs
Publication statusPublished - Mar 2022

Fingerprint

Dive into the research topics of 'The reliability of CT numbers as absolute values for diagnostic scanning, dental imaging, and radiation therapy simulation: A narrative review'. Together they form a unique fingerprint.

Cite this