Using aluminum for scatter control in mammography: preliminary work using measurements of CNR and FOM

Khaled Al Khalifah, Rob Davidson, Abel Zhou

Research output: Contribution to journalArticle

Abstract

Full-field digital mammography (FFDM) systems provide the current gold standard in mammographic examinations. Although FFDM provides the lowest mammographic doses, the radiation dose to the breast during mammographic examinations is still a concern. Thus, image quality optimization at the lowest dose is a major goal. In planar X-ray imaging, thin sheets of aluminum (Al) are used as filtration to reduce the number of low-energy X-ray photons reaching the patient. The goal of this work was to evaluate whether Al can be used in FFDM to remove scatter radiation from reaching the image detector, hence improving image quality. Doses were compared with the use of a grid. A Hologic Selenia mammographic unit was used to acquire images of two phantoms, namely, the ACR phantom and a Perspex phantom of 5 cm. Images were acquired using two tube voltages (kVp) and filter combinations under two exposure/dose conditions. Al sheets of various thicknesses were placed between the phantom and the image detector. Contrast-to-noise ratio (CNR) and figure of merit (FOM) values were measured and compared with images acquired using a grid. When a constant dose was delivered to the image detector, the highest CNR was achieved using a grid; however, the highest FOM values were achieved when using 0.05 mm thick Al sheets. This study successfully demonstrates that thin sheets of Al can be used in mammography examinations to reduce scattered radiation and improve image quality, as indicated by the measured CNR values. Given the limitations of this work, further kVp and target/filter combinations and various methods of image quality measurement need to be studied.

Original languageEnglish
Pages (from-to)1-10
Number of pages10
JournalRadiological Physics and Technology
Early online date20 Nov 2019
DOIs
Publication statusPublished - 20 Nov 2019

Fingerprint

Mammography
Aluminum
figure of merit
Noise
aluminum
Radiation
dosage
X-Rays
examination
grids
Polymethyl Methacrylate
Selenium
Photons
detectors
Breast
radiation
Perspex (trademark)
filters
digital systems
selenium

Cite this

@article{2a58cafd996b402683c46eb9e02cc491,
title = "Using aluminum for scatter control in mammography: preliminary work using measurements of CNR and FOM",
abstract = "Full-field digital mammography (FFDM) systems provide the current gold standard in mammographic examinations. Although FFDM provides the lowest mammographic doses, the radiation dose to the breast during mammographic examinations is still a concern. Thus, image quality optimization at the lowest dose is a major goal. In planar X-ray imaging, thin sheets of aluminum (Al) are used as filtration to reduce the number of low-energy X-ray photons reaching the patient. The goal of this work was to evaluate whether Al can be used in FFDM to remove scatter radiation from reaching the image detector, hence improving image quality. Doses were compared with the use of a grid. A Hologic Selenia mammographic unit was used to acquire images of two phantoms, namely, the ACR phantom and a Perspex phantom of 5 cm. Images were acquired using two tube voltages (kVp) and filter combinations under two exposure/dose conditions. Al sheets of various thicknesses were placed between the phantom and the image detector. Contrast-to-noise ratio (CNR) and figure of merit (FOM) values were measured and compared with images acquired using a grid. When a constant dose was delivered to the image detector, the highest CNR was achieved using a grid; however, the highest FOM values were achieved when using 0.05 mm thick Al sheets. This study successfully demonstrates that thin sheets of Al can be used in mammography examinations to reduce scattered radiation and improve image quality, as indicated by the measured CNR values. Given the limitations of this work, further kVp and target/filter combinations and various methods of image quality measurement need to be studied.",
keywords = "Aluminum sheets, CNR, FOM, Full-field digital mammography, Image quality, Radiation dose",
author = "{Khaled Al Khalifah} and Rob Davidson and Abel Zhou",
year = "2019",
month = "11",
day = "20",
doi = "10.1007/s12194-019-00545-3",
language = "English",
pages = "1--10",
journal = "Radiological Physics and Technology",
issn = "1865-0333",
publisher = "Springer",

}

TY - JOUR

T1 - Using aluminum for scatter control in mammography: preliminary work using measurements of CNR and FOM

AU - Khaled Al Khalifah

AU - Davidson, Rob

AU - Zhou, Abel

PY - 2019/11/20

Y1 - 2019/11/20

N2 - Full-field digital mammography (FFDM) systems provide the current gold standard in mammographic examinations. Although FFDM provides the lowest mammographic doses, the radiation dose to the breast during mammographic examinations is still a concern. Thus, image quality optimization at the lowest dose is a major goal. In planar X-ray imaging, thin sheets of aluminum (Al) are used as filtration to reduce the number of low-energy X-ray photons reaching the patient. The goal of this work was to evaluate whether Al can be used in FFDM to remove scatter radiation from reaching the image detector, hence improving image quality. Doses were compared with the use of a grid. A Hologic Selenia mammographic unit was used to acquire images of two phantoms, namely, the ACR phantom and a Perspex phantom of 5 cm. Images were acquired using two tube voltages (kVp) and filter combinations under two exposure/dose conditions. Al sheets of various thicknesses were placed between the phantom and the image detector. Contrast-to-noise ratio (CNR) and figure of merit (FOM) values were measured and compared with images acquired using a grid. When a constant dose was delivered to the image detector, the highest CNR was achieved using a grid; however, the highest FOM values were achieved when using 0.05 mm thick Al sheets. This study successfully demonstrates that thin sheets of Al can be used in mammography examinations to reduce scattered radiation and improve image quality, as indicated by the measured CNR values. Given the limitations of this work, further kVp and target/filter combinations and various methods of image quality measurement need to be studied.

AB - Full-field digital mammography (FFDM) systems provide the current gold standard in mammographic examinations. Although FFDM provides the lowest mammographic doses, the radiation dose to the breast during mammographic examinations is still a concern. Thus, image quality optimization at the lowest dose is a major goal. In planar X-ray imaging, thin sheets of aluminum (Al) are used as filtration to reduce the number of low-energy X-ray photons reaching the patient. The goal of this work was to evaluate whether Al can be used in FFDM to remove scatter radiation from reaching the image detector, hence improving image quality. Doses were compared with the use of a grid. A Hologic Selenia mammographic unit was used to acquire images of two phantoms, namely, the ACR phantom and a Perspex phantom of 5 cm. Images were acquired using two tube voltages (kVp) and filter combinations under two exposure/dose conditions. Al sheets of various thicknesses were placed between the phantom and the image detector. Contrast-to-noise ratio (CNR) and figure of merit (FOM) values were measured and compared with images acquired using a grid. When a constant dose was delivered to the image detector, the highest CNR was achieved using a grid; however, the highest FOM values were achieved when using 0.05 mm thick Al sheets. This study successfully demonstrates that thin sheets of Al can be used in mammography examinations to reduce scattered radiation and improve image quality, as indicated by the measured CNR values. Given the limitations of this work, further kVp and target/filter combinations and various methods of image quality measurement need to be studied.

KW - Aluminum sheets

KW - CNR

KW - FOM

KW - Full-field digital mammography

KW - Image quality

KW - Radiation dose

UR - http://www.scopus.com/inward/record.url?scp=85075355048&partnerID=8YFLogxK

U2 - 10.1007/s12194-019-00545-3

DO - 10.1007/s12194-019-00545-3

M3 - Article

SP - 1

EP - 10

JO - Radiological Physics and Technology

JF - Radiological Physics and Technology

SN - 1865-0333

ER -